Skip to product information
1 of 1
CLIA Kit for Coagulation Factor VIII (F8) TRV-053746-01

CLIA Kit for Coagulation Factor VIII (F8)

Size

SKU:TRV-053746-01

Regular price $308.39
Regular price Sale price $308.39
Sale Sold out
Shipping calculated at checkout.
Quantity

CLIA Kit for Coagulation Factor VIII (F8) is a premium-quality life science research product supplied by Torvigen for laboratory and scientific applications.

View full details

Product specifications

Alternative Name FVIII; AHF; HEMA; F8B; F8C; Antihemophilic Globulin; Hemophilia A; Anti-Hemophilic Factor A; FVIII Procoagulant Component; F8 Protein
Name Synonym Coagulation Factor VIII
Uniprot P00741
Field of Research Metabolic pathway; Hematology
Applications Chemiluminescent immunoassay for Antigen Detection.
Organism Species Homo sapiens (Human)
Format 48T, 96T, 96Tx5, 96Tx10, 96Tx100
Specificity This assay has high sensitivity and excellent specificity for detection of Coagulation Factor VIII (F8) . No significant cross-reactivity or interference between Coagulation Factor VIII (F8) and analogues was observed.
Sensitivity The minimum detectable dose of this kit is typically less than 4.9pg/mL
Detection Range 13.7-10,000pg/mL
Precision Intra-assay Precision (Precision within an assay) : 3 samples with low, middle and high level Coagulation Factor VIII (F8) were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays) : 3 samples with low, middle and high level Coagulation Factor VIII (F8) were tested on 3 different plates, 8 replicates in each plate. CV (%) = SD/meanX100 Intra-Assay: CV Inter-Assay: CV Inter-Assay: CV
Stability The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Assay Principle The microplate provided in this kit has been pre-coated with an antibody specific to Coagulation Factor VIII (F8) . Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Coagulation Factor VIII (F8) . Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Coagulation Factor VIII (F8) level in the sample or standard.
Assay Protocol 1. Prepare all reagents, samples and standards;
2. Add 100µ L standard or sample to each well. Incubate 2 hours at 37° C;
3. Aspirate and add 100µ L prepared Detection Reagent A. Incubate 1 hour at 37° C;
4. Aspirate and wash 3 times;
5. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C;
6. Aspirate and wash 5 times;
7. Add 100µ L Substrate Solution. Incubate 10 minutes at 37° C;
8. Read RLU value immediately.
Assay Performance Time 2h, 40min
Method Double-antibody Sandwich
Sample Type Plasma